Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF
Zeng, JR; Bian, FG; Wang, J; Li, XH; Wang, YZ; Tian, F; Zhou, P
2017
发表期刊JOURNAL OF SYNCHROTRON RADIATION
ISSN1600-5775
卷号24页码:509-520
文章类型期刊论文
摘要The optical system and end-station of bending-magnet beamline BL16B1, dedicated to small-angle X-ray scattering (SAXS) at the Shanghai Synchrotron Radiation Facility, is described. Constructed in 2009 and upgraded in 2013, this beamline has been open to users since May 2009 and supports methodologies including SAXS, wide-angle X-ray scattering (WAXS), simultaneous SAXS/WAXS, grazing-incidence small-angle X-ray scattering (GISAXS) and anomalous small-angle X-ray scattering (ASAXS). Considering that an increasing necessity for absolute calibration of SAXS intensity has been recognized in in-depth investigations, SAXS intensity is re-stated according to the extent of data processing, and the absolute intensity is suggested to be a unified presentation of SAXS data in this article. Theory with a practical procedure for absolute intensity calibration is established based on BL16B1, using glass carbon and water as primary and secondary standards, respectively. The calibration procedure can be completed in minutes and shows good reliability under different conditions. An empirical line of scale factor estimation is also established for any specific SAXS setup at the beamline. Beamline performance on molecular weight (MW) determination is provided as a straightforward application and verification of the absolute intensity calibration. Results show good accuracy with a deviation of less than 10% compared with the known value, which is also the best attainable accuracy in recent studies using SAXS to measure protein MW. Fast MW measurement following the demonstrated method also enables an instant check or pre-diagnosis of the SAXS performance to improve the data acquisition.
关键词Beamline Saxs Absolute Intensity Calibration Molecular Weight
DOI10.1107/S1600577516019135
收录类别SCI
语种英语
WOS记录号WOS:000395093600019
引用统计
文献类型期刊论文
条目标识符http://ir.sinap.ac.cn/handle/331007/27409
专题中科院上海应用物理研究所2011-2018年
推荐引用方式
GB/T 7714
Zeng, JR,Bian, FG,Wang, J,et al. Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF[J]. JOURNAL OF SYNCHROTRON RADIATION,2017,24:509-520.
APA Zeng, JR.,Bian, FG.,Wang, J.,Li, XH.,Wang, YZ.,...&Zhou, P.(2017).Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF.JOURNAL OF SYNCHROTRON RADIATION,24,509-520.
MLA Zeng, JR,et al."Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF".JOURNAL OF SYNCHROTRON RADIATION 24(2017):509-520.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Performance on absol(1322KB)期刊论文作者接受稿开放获取CC BY-NC-SA请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Zeng, JR]的文章
[Bian, FG]的文章
[Wang, J]的文章
百度学术
百度学术中相似的文章
[Zeng, JR]的文章
[Bian, FG]的文章
[Wang, J]的文章
必应学术
必应学术中相似的文章
[Zeng, JR]的文章
[Bian, FG]的文章
[Wang, J]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。